DSpace Repository

A fieldable, high-throughput, cost-efficient high performance liquid chromatography-ultraviolet absorption detection (HPLC-UV) method for the quantitation of bispyridinium quaternary aldoxime cholinesterase reactivators in blood

Show simple item record

dc.rights.license CC BY eng
dc.contributor.author Karvaly, Gellert Balazs cze
dc.contributor.author Tekes, Kornelia cze
dc.contributor.author Szimrok, Zoltan cze
dc.contributor.author Furesz, Jozsef cze
dc.contributor.author Kuča, Kamil cze
dc.contributor.author Kalasz, Huba cze
dc.date.accessioned 2026-07-21T06:19:44Z
dc.date.available 2026-07-21T06:19:44Z
dc.date.issued 2021 eng
dc.identifier.issn 1233-2356 eng
dc.identifier.uri http://hdl.handle.net/20.500.12603/2746
dc.description.abstract Mono- and bis-pyridinium quaternary aldoximes (K-oximes) have long been employed as cholinesterase reactivator components of antidotes against lethal cholinesterase-inhibiting organophosphorous chemicals. Their positive charge poses difficulties in their chromatographic analysis, resulting in the publication of different approaches for each K-oxime. A multiplexed method is presented for the rapid quantitation of 10 K-oximes in blood with its utility demonstrated in vivo. Liquid chromatography with absorbance detection was employed. Reversed-phase separation was achieved on a highly nonpolar stationary phase. Method validation was based on the respective guideline of the European Medicines Agency. Times to peak concentrations and 120-min areas under the time-concentration curves were determined in rats following intraperitoneal administration. Adequate retention and separation of K-oximes with acceptable peak shapes in short isocratic runs was achieved by adjusting ionic strength, organic content and the concentration of the ion-pairing agent of the mobile phase. Chromatographic properties were governed by optimizing the concentration of dissolved ions. Accurate adjustment of the organic content was indispensable for avoiding peak drifting and splitting. Dose-adjusted exposure to K-347 and K-868 was exceptionally low, while exposure to K-48 was the highest. The method is suitable for screening systemic exposure to various K-oximes and can be extended. eng
dc.format p. 134-144 eng
dc.language.iso eng eng
dc.publisher AKADEMIAI KIADO ZRT eng
dc.relation.ispartof ACTA CHROMATOGRAPHICA, volume 33, issue: 2 eng
dc.subject pralidoxime eng
dc.subject obidoxime eng
dc.subject cholinesterase eng
dc.subject cholinesterase reactivator eng
dc.subject nerve agent eng
dc.title A fieldable, high-throughput, cost-efficient high performance liquid chromatography-ultraviolet absorption detection (HPLC-UV) method for the quantitation of bispyridinium quaternary aldoxime cholinesterase reactivators in blood eng
dc.type article eng
dc.identifier.obd 43877615 eng
dc.identifier.wos 000631178700006 eng
dc.identifier.doi 10.1556/1326.2020.00781 eng
dc.publicationstatus postprint eng
dc.peerreviewed yes eng
dc.source.url https://akjournals.com/view/journals/1326/aop/article-10.1556-1326.2020.00781/article-10.1556-1326.2020.00781.xml cze
dc.relation.publisherversion https://akjournals.com/view/journals/1326/aop/article-10.1556-1326.2020.00781/article-10.1556-1326.2020.00781.xml eng
dc.rights.access Open Access eng


Files in this item